Load time: 03/19/2012 12:25:35 PM Read time: 03/19/2012 12:25:35 PM Kinetic read cycle: 1 of 1 Read start delay time: 0 s Time delay between kinetic reads: Barcode: ERROR Method ID: Default Fluorescence Intensity Method Plate ID: #2338 Comment: new comment Microplate format: Corning Costar Clear Bottom PS Shake time: 0 s Temperature: 30.9 C Instrument tag: Serial number: AD2123 Read sequence: column Mode sequence: plate Detection mode: f0 Light sensor: HC-120 Excitation side: Top Emission side: Top Lamp: Continuous Readings per well: 1 Time between readings: Integration time: 100000 us Attenuator mode: m Motion settling time: 25 ms Z Height: 3 mm Numeric Excitation filter: 1 Fluorescein 485 nm Emission filter: 1 Fluorescein 530 nm Beamsplitter: Top Fluorescein 505 Excitation polarizer filter: o Emission polarizer filter: o Detector counting: SmartRead Flash lamp voltage: Delay after flash: Sensitivity setting: 4 A/D converter gain: x1 Integrating gain: x1 Max cps: 203449 cps Min counts: 3448 counts Well list: r0(a1:e12) = (f0(a1:e12)) Data: RAW DATA Units: rfu Display format: %.2f 1 2 3 4 5 6 7 8 9 10 11 12 A 18080400.00 14817400.00 13115700.00 11732000.00 11357800.00 11331300.00 11292700.00 11416500.00 11189600.00 10891700.00 11007800.00 10596400.00 B 20344900.00 16035000.00 13449300.00 12073600.00 12110800.00 11048900.00 14196600.00 10827500.00 10941700.00 11472100.00 10952700.00 10910800.00 C 14013300.00 12396100.00 12194600.00 11937600.00 11717500.00 11057300.00 11065600.00 11225600.00 11514400.00 11483500.00 10782800.00 18095600.00 D 17628800.00 12934000.00 12170000.00 11399200.00 11908100.00 11111700.00 10861500.00 11193400.00 11700400.00 11272500.00 10893300.00 10480200.00 E 5871400.00 5818500.00 5982600.00 5707100.00 5753900.00 5357500.00 5484200.00 5622000.00 5504200.00 5693200.00 5581500.00 5439900.00 F G H Load time: 03/19/2012 12:28:57 PM Read time: 03/19/2012 12:28:57 PM Kinetic read cycle: 1 of 1 Read start delay time: 0 s Time delay between kinetic reads: Barcode: ERROR Method ID: DNATech Fluorescence Int Plate ID: #2339 Comment: read 2 Microplate format: MJRBlackSkirt96 Shake time: 0 s Temperature: 30.9 C Instrument tag: Serial number: AD2123 Read sequence: row Mode sequence: plate Detection mode: f1 Light sensor: HC-120 Excitation side: Top Emission side: Top Lamp: Continuous Readings per well: 1 Time between readings: Integration time: 100000 us Attenuator mode: m Motion settling time: 25 ms Z Height: 3 mm Numeric Excitation filter: 1 Fluorescein 485 nm Emission filter: 1 Fluorescein 530 nm Beamsplitter: Top Fluorescein 505 Excitation polarizer filter: o Emission polarizer filter: o Detector counting: SmartRead Flash lamp voltage: Delay after flash: Sensitivity setting: 4 A/D converter gain: x1 Integrating gain: x1 Max cps: 194524 cps Min counts: 3448 counts Well list: r0(a1:a12,b1:b12,c1:c12,d1:d12,e1:e12,f1:f12,g1:g12,h1:h12) = (f1(a1:a12,b1:b12,c1:c12,d1:d12,e1:e12,f1:f12,g1:g12,h1:h12)) Data: RAW DATA Units: rfu Display format: %.2f 1 2 3 4 5 6 7 8 9 10 11 12 A 17280300.00 16193900.00 14572900.00 13988200.00 13038300.00 12926800.00 12682000.00 12567700.00 12347900.00 12805900.00 12475100.00 11954300.00 B 19452400.00 16109200.00 14708000.00 14140000.00 13333700.00 13104700.00 12888700.00 12580000.00 12806700.00 12616800.00 12851700.00 12418700.00 C 14765900.00 13945400.00 13177400.00 12533100.00 13708800.00 13230800.00 13162300.00 12944000.00 12802100.00 12633000.00 12553300.00 12525900.00 D 17542400.00 13913500.00 13475900.00 13126900.00 13026000.00 11858300.00 12654900.00 12371000.00 12811700.00 12601100.00 12691400.00 12221700.00 E 8807100.00 8689500.00 8979700.00 8943400.00 9095800.00 8592600.00 8589800.00 8759800.00 9056600.00 8771000.00 8717100.00 9054400.00 F 8980100.00 8838700.00 9312000.00 9217100.00 9395900.00 10230600.00 9081000.00 9192000.00 9082100.00 9065900.00 8727900.00 8378200.00 G 8932400.00 13104900.00 9237800.00 9031600.00 9089500.00 9120400.00 9221300.00 9379800.00 8872400.00 9003300.00 9200900.00 8616600.00 H 8784800.00 9125300.00 8981200.00 8828600.00 9369800.00 9251400.00 8800700.00 9233800.00 9015900.00 9191600.00 8704400.00 8424300.00 Load time: 03/19/2012 12:30:23 PM Read time: 03/19/2012 12:30:23 PM Kinetic read cycle: 1 of 1 Read start delay time: 0 s Time delay between kinetic reads: Barcode: ERROR Method ID: DNATech Fluorescence Int Plate ID: #2340 Comment: read 2 Microplate format: MJRBlackSkirt96 Shake time: 0 s Temperature: 31 C Instrument tag: Serial number: AD2123 Read sequence: row Mode sequence: plate Detection mode: f1 Light sensor: HC-120 Excitation side: Top Emission side: Top Lamp: Continuous Readings per well: 1 Time between readings: Integration time: 100000 us Attenuator mode: m Motion settling time: 25 ms Z Height: 3 mm Numeric Excitation filter: 1 Fluorescein 485 nm Emission filter: 1 Fluorescein 530 nm Beamsplitter: Top Fluorescein 505 Excitation polarizer filter: o Emission polarizer filter: o Detector counting: SmartRead Flash lamp voltage: Delay after flash: Sensitivity setting: 4 A/D converter gain: x1 Integrating gain: x1 Max cps: 198581 cps Min counts: 3448 counts Well list: r0(a1:d12) = (f1(a1:d12)) Data: RAW DATA Units: rfu Display format: %.2f 1 2 3 4 5 6 7 8 9 10 11 12 A 17271100.00 15232800.00 14155500.00 13206700.00 13072000.00 12410100.00 12359200.00 12941400.00 12238100.00 12062700.00 12747700.00 12041600.00 B 19858100.00 16595400.00 15232700.00 13671200.00 13562500.00 12914500.00 12749900.00 12719600.00 12764700.00 12418700.00 12444700.00 12173500.00 C 14727800.00 13912800.00 14167200.00 12832000.00 13582800.00 12721800.00 12813300.00 12190400.00 12645300.00 12720300.00 12429800.00 12552600.00 D 18058300.00 13878000.00 13186000.00 12922200.00 12437000.00 12574800.00 12175500.00 12608100.00 12954000.00 12564100.00 12214300.00 12061600.00 E F G H Load time: 03/19/2012 12:32:56 PM Read time: 03/19/2012 12:32:56 PM Kinetic read cycle: 1 of 1 Read start delay time: 0 s Time delay between kinetic reads: Barcode: ERROR Method ID: DNATech Fluorescence Int Plate ID: #2341 Comment: samples 1 Microplate format: MJRBlackSkirt96 Shake time: 0 s Temperature: 31.1 C Instrument tag: Serial number: AD2123 Read sequence: row Mode sequence: plate Detection mode: f1 Light sensor: HC-120 Excitation side: Top Emission side: Top Lamp: Continuous Readings per well: 1 Time between readings: Integration time: 100000 us Attenuator mode: m Motion settling time: 25 ms Z Height: 3 mm Numeric Excitation filter: 1 Fluorescein 485 nm Emission filter: 1 Fluorescein 530 nm Beamsplitter: Top Fluorescein 505 Excitation polarizer filter: o Emission polarizer filter: o Detector counting: SmartRead Flash lamp voltage: Delay after flash: Sensitivity setting: 4 A/D converter gain: x1 Integrating gain: x1 Max cps: 2568367 cps Min counts: 3448 counts Well list: r0(a1:g12) = (f1(a1:g12)) Data: RAW DATA Units: rfu Display format: %.2f 1 2 3 4 5 6 7 8 9 10 11 12 A 57215400.00 129513100.00 110721900.00 133127900.00 122081000.00 13074100.00 92782600.00 49660800.00 78601100.00 105744700.00 122811500.00 104122700.00 B 124716900.00 86453500.00 104698100.00 94881000.00 118113000.00 20034600.00 18052200.00 111675700.00 64457600.00 88123700.00 117365300.00 129503400.00 C 74838200.00 113237800.00 142755500.00 93231600.00 53613800.00 32648600.00 18893100.00 103448100.00 15708100.00 15005400.00 19514100.00 165033800.00 D 131448300.00 83584900.00 160092200.00 135438900.00 17147500.00 14705000.00 14477500.00 97667600.00 28428900.00 13618100.00 89059100.00 232569900.00 E 92862600.00 134205500.00 109390800.00 10472800.00 15380300.00 14295900.00 15607400.00 126519600.00 14279100.00 96549200.00 128753900.00 118894900.00 F 113483800.00 113476300.00 95051000.00 71360000.00 105489100.00 147879300.00 91631300.00 72914100.00 146013400.00 131138400.00 114182700.00 12232100.00 G 125292300.00 135780200.00 109616800.00 101757000.00 83520100.00 69778900.00 20850700.00 56964000.00 256836700.00 202641100.00 196442400.00 252622200.00 H Load time: 03/19/2012 12:34:47 PM Read time: 03/19/2012 12:34:47 PM Kinetic read cycle: 1 of 1 Read start delay time: 0 s Time delay between kinetic reads: Barcode: ERROR Method ID: DNATech Fluorescence Int Plate ID: #2342 Comment: different read depth Microplate format: MJRBlackSkirt96 Shake time: 0 s Temperature: 31.1 C Instrument tag: Serial number: AD2123 Read sequence: row Mode sequence: plate Detection mode: f1 Light sensor: HC-120 Excitation side: Top Emission side: Top Lamp: Continuous Readings per well: 1 Time between readings: Integration time: 100000 us Attenuator mode: m Motion settling time: 25 ms Z Height: 0 mm Bottom Excitation filter: 1 Fluorescein 485 nm Emission filter: 1 Fluorescein 530 nm Beamsplitter: Top Fluorescein 505 Excitation polarizer filter: o Emission polarizer filter: o Detector counting: SmartRead Flash lamp voltage: Delay after flash: Sensitivity setting: 4 A/D converter gain: x1 Integrating gain: x1 Max cps: 1298154 cps Min counts: 3448 counts Well list: r0(a1:g12) = (f1(a1:g12)) Data: RAW DATA Units: rfu Display format: %.2f 1 2 3 4 5 6 7 8 9 10 11 12 A 31067700.00 63546600.00 59536700.00 61317300.00 63993500.00 8216800.00 50959700.00 26895700.00 43591200.00 61540500.00 64916200.00 52322800.00 B 66114700.00 44368500.00 55018200.00 53069200.00 65952900.00 11828800.00 11048400.00 56729700.00 35341800.00 47171800.00 58893100.00 65132000.00 C 39926600.00 57700800.00 76249700.00 47703000.00 30089900.00 17568800.00 10375100.00 53842300.00 9795700.00 8254600.00 11677700.00 84876800.00 D 66840000.00 43538400.00 86823600.00 77323400.00 10128800.00 8920000.00 8593900.00 53472400.00 15435800.00 8568700.00 49891400.00 115255000.00 E 52903900.00 69141000.00 54238600.00 8199500.00 8834400.00 9187300.00 9022200.00 72264700.00 8656500.00 50268100.00 69869400.00 63415700.00 F 55350300.00 60658300.00 53326900.00 38453400.00 56551100.00 74864200.00 52458100.00 38609300.00 76810100.00 84918600.00 58157100.00 7080500.00 G 69804300.00 73161600.00 58593900.00 56260500.00 44785700.00 36070700.00 12066100.00 29398600.00 129815400.00 109927800.00 107605600.00 124012700.00 H